2.4.2 Types of Instrumentation
There
are basically two types of sensors for measuring humidity:
- psychrometers
- and
hygrometers.
The psychrometer, consists of two thermometers, one of which is
covered with a wet wick (the wet bulb) and a mechanism for ventilating the
pair. Evaporation lowers the temperature of the wet bulb; the difference in
temperature from the dry bulb (the wet bulb depression) is a measure of the
amount of moisture in the air. While still in use at many observing
stations, psychrometers are generally not suitable for routine monitoring
programs. However, they can be used as secondary standards in audit
procedures.
Hygrometers
are a class of instruments that measure the physical effect that moisture
has on a substances, such as hair. For example, the lithium chloride
hygrometer uses a probe impregnated with lithium chloride solution. Voltage
is supplied to the electrodes in the probe until an equilibrium temperature
is reached based on the conductivity of the lithium chloride.
The dew point
hygrometer, uses a cooled mirror as a sensor; in this case, the temperature
of the mirror is monitored to determine the temperature at which dew (or
frost) first appears. Such condensation typically disrupts the path of a
light beam reflecting off of the cooled surface, causing it to be heated
until the condensation disappears. Once the condensation is gone, the
surface is cooled again until condensation forms. These oscillating heating
and cooling cycles define an average dew point temperature. The temperature
of the surface is typically measured by a linear thermistor or a platinum
RTD. The thin film capacitor hygrometer measures humidity by detecting the
change in capacitance of a thin polymer film; this sensor has a relatively
fast response compared to other types of hygrometers.
If
possible, humidity sensors should be housed in the same aspirated radiation
shield as the temperature sensor. The humidity sensor should be protected
from contaminants such as salt, hydrocarbons, and other particulates. The
best protection is the use of a porous membrane filter which allows the
passage of ambient air and water vapor while keeping out particulate matter.
2. PRIMARY METEOROLOGICAL VARIABLES
2.1 Wind Speed
2.1.1 Cup Anemometers
2.1.2 Vane-oriented and Fixed-mount Propeller Anemometers
2.1.3 Wind Speed Transducers
2.2 Wind Direction
2.2.1 Wind Vanes
2.2.2 U-V and UVW Systems
2.2.3 Wind Direction Transducers
2.2.4 Standard Deviation and Turbulence
Data
2.3 Temperature and Temperature Difference
2.3.1 Classes of Temperature Sensors
2.3.2 Response Characteristics
2.3.3 Temperature Difference
2.3.4 Sources of Error
2.4 Humidity
2.4.1 Humidity Variables
2.4.2 Types of Instrumentation
2.5 Precipitation
2.6 Pressure
2.7 Radiation
2.8 Recommendations