8.3.1 Sensor Check
There
are three types of action which can be considered a sensor check. First, one
can look at and perform "housekeeping" services for the sensors.
Secondly, one can measure some attribute of the sensor to detect
deterioration in anticipation of preventative maintenance. Thirdly, the
sensor can be subjected to a known condition whose consequence Is
predictable through the entire measurement system, including the sensor
transducer. Each of these will be addressed for each variable, where
appropriate, within the divisions of physical inspection and measurement and
accuracy check with known input.
Physical
inspection
The
first level of inspection is visual. The anemometer and vane can be looked
at, either directly or through binoculars or a telescope, to check for
physical damage or signs of erratic behavior. Temperature shields can
be checked for cleanliness. Precipitation gauges can be inspected for
foreign matter which might effect performance. The static port for the
atmospheric pressure system also can be examined for foreign matter. Solar
radiation sensors should be wiped clean at every opportunity.
A
better level of physical inspection is a "hands on" check. An
experienced techniciancan feel the condition of the anemometer bearing
assembly and know whether or not they are in good condition. This is best
done with the aerodynamic shape (cup wheel, propeller, or vane) removed.
Caution: Damage to anemometers and vanes is more likely to result from human
handling than from the forces of the wind, especially during removal or
installation and transport up and down a tower. The proper level of
aspiration through a forced aspiration shield can be felt and heard under
calm condition.
The
best level of sensor check is a measurement. The anemometer and wind vane
sensors have bearings which will certainly degrade in time. The goal is to
change the bearings or the sensors before the instrument falls below
operating specifications. Measurements of starting torque will provide the
objective data upon which maintenance decisions can be made and defended.
The presence, in routine calibration reports, of starting torque
measurements will support the claim for valid data, if the values are less
than the replacement torques.
The
anemometer, identified by the serial number of the aerodynamic shape, should
have a wind tunnel calibration report (see Section 8.1) in a permanent
record folder. This is the authority for the transfer function (rate of
rotation to wind speed) to be used in the next section. The temperature
transducers, identified by serial number, should have calibration reports
showing their conformity for at least three points to their generic transfer
function (resistance to temperature, usually). These reports should specify
the instruments used for the calibration and the method by which the
instruments are tied to national standards (NBS). The less important sensors
for solar radiation and atmospheric pressure can be qualified during an
audit for accuracy.
Accuracy
check with known input
Two
simple tests will determine the condition of the anemometer (assuming no
damage is found by the physical inspection). The aerodynamic shape must be
removed. The shaft is driven at three known rates of rotation. The rates are
known by independently counting shaft revolutions over a measured period of
time in synchronization with the measurement system timing. The rates should
be meaningful such as the equivalent of 2 m/s, 5 m/s and 10 m/s. Conversion
of rates of rotation to wind speed is done with the manufacturer's transfer
function or wind tunnel data. For example, if the transfer function is m/s =
1.412 r/s + 0.223, then rates of rotation of 1.3, 3.4 and 6.9 revolutions
per second (r/s) would be equivalent to about 2, 5 and 10 m/s. All that is
being tested is the implementation of the transfer function by the measuring
system. The output should agree within one increment of resolution (probably
0.1 m/s). If problems are found, they might be in the transducer, although
failures there are usually catastrophic. The likely source of trouble is the
measurement system (signal conditioner, transmitting system, averaging
system and recording system).
The
second test is for starting torque. This test requires a torque watch or
similar device capable of measuring in the range of 0.1 to 10 gm-cm
depending upon the specifications provided by the manufacturer.
A
successful response to these two tests will document the fact that the
anemometer is operating as well as it did
at receiving inspection, having verified threshold and accuracy.
Changes
in distance constant are not likely unless the anemometer design has
changed. If a plastic cup is replaced by a stainless steel cup, for example,
both the transfer function and the distance constant will likely be
different. The distance constant will vary as the inverse of the air
density. If a sea-level distance constant is 3.0 m, it may increase to 3.5 m
in Denver and 4.3 m at the mountain passes in the Rockies.
For
wind direction, a fixture holding the vane, or vane substitute, in positions
with a known angle change is a fundamental challenge to the relative
accuracy of the wind vane. With this method, applying the appropriate
strategy for 360 or 540 degree systems, the accuracy of the sensor can be
documented. The accuracy of the wind direction measurement, however, also
depends on the orientation of the sensor with respect to true north.
The
bearing to distant objects may be determined by several methods. The
recommended method employs a solar observation (see Reference 3, p.11) to
find the true north-south line where it passes through the sensor mounting
location. Simple azimuth sighting devices can be used to find the bearing of
some distant object with respect to the north-south line. The "as
found" and "as left" orientation readings should report the
direction to or from that distant object. The object should be one toward
which the vane can be easily aimed and not likely to become hidden by
vegetation or construction.
There
are two parts of most direction vanes which wear out. One part is the
bearing assembly and the other is the transducer, usually a potentiometer.
Both contribute to the starting torque and hence the threshold of the
sensor. A starting torque measurement will document the degradation of the
threshold and flag the need for preventive maintenance. An analog voltmeter
or oscilloscope is required to see the noise level of a potentiometer.
Transducer noise may not be a serious problem with average values but it is
likely to have a profound effect on
A
.
The
dynamic performance characteristics of a wind vane are best measured with a
wind tunnel test. A generic test of a design sample is adequate. As with the
anemometer, the dynamic response characteristics (threshold, delay distance
and damping ratio) are density dependent.
Temperature
transducers are reasonably stable, but they may drift with time. The known
input for a temperature transducer is a stable thermal mass whose
temperature is known by a standard transducer. The ideal thermal mass is one
with a time constant on the order of an hour in which there are no thermal
sources or sinks to establish local gradients within the mass. It is far
more important to know what a mass temperature is than to be able to set a
mass to a particular temperature.
For
temperature difference systems, the immersion of all transducers in a single
mass as described above will provide a zero-difference challenge accurate to
about 0.01 °C. When this test is repeated with the mass at two more
temperatures, the transducers will have been challenged with respect to how
well they are matched and how well they follow the generic. Transfer
function. Mass temperatures in the ranges of 0 to 10 °C, 15 to 25 °C, and
30 to 40 °C are recommended. A maximum difference among the three
temperatures (i.e., 0, 20, and 40 °C) is optimum. Once the match has been
verified, known resistances can be substituted for the transducers
representing temperatures, according to the generic transfer function,
selected to produce known temperature difference signals to the signal
conditioning circuitry. This known input will challenge the circuitry for
the differential measurement.
Precipitation
sensors can be challenged by inserting a measured amount of water, at
various reasonable rainfall rates such as 25 mm or less per hour. The area
of the collector can be measured to calculate the amount of equivalent
rainfall which was inserted. The total challenge should be sufficient to
verify a 10% accuracy in measurement of water. This does not provide
information about errors from siting problems or wind effects.
Dew
point temperature (or relative humidity), atmospheric pressure and radiation
are most simply challenged in an ambient condition with a collocated
transfer standard. An Assmann psychrometer may be used for dew point. An
aneroid barometer checked against a local National Weather Service
instrument is recommended for atmospheric pressure. Another radiation sensor
with some pedigree or manufacturer's certification may be used for
pyranometers and net radiometers. A complete opaque cover will provide a
zero check.
8. QUALITY ASSURANCE AND QUALITY CONTROL
8.1 Instrument Procurement
8.1.1 Wind Speed
8.1.2 Wind Direction
8.1.3 Temperature and Temperature Difference
8.1.4 Dew Point Temperature
8.1.5 Precipitation
8.1.6 Pressure
8.1.7 Radiation
8.2 Installation and Acceptance Testing
8.2.1 Wind Speed
8.2.2 Wind Direction
8.2.3 Temperature and Temperature Difference
8.2.4 Dew Point Temperature
8.2.5 Precipitation
8.2.6 Pressure
8.2.7 Radiation
8.3 Routine Calibrations
8.3.1 Sensor Check
8.3.2 Signal Conditioner and Recorder Check
8.3.3 Calibration Data Logs
8.3.4 Calibration Report
8.3.5 Calibration Schedule/Frequency
8.3.6 Data Correction Based on Calibration Results
8.4 Audits
8.4.1 Audit Schedule and Frequency
8.4.2 Audit Procedure
8.4.3 Corrective Action and Reporting
8.5 Routine and Preventive Maintenance
8.5.1 Standard Operating Procedures
8.5.2 Preventive Maintenance
8.6 Data Validation and Reporting
8.6.1 Preparatory Steps
8.6.2 Levels of Validation
8.6.3 Validation Procedures
8.6.4 Schedule and Reporting
8.7 Recommendations