8.3.2 Signal Conditioner and Recorder Check
For
routine calibration of measurement circuits and recorders, use the
manufacturer's recommendations. The outputs required by the test described
in 8.3.1.2 must be reflected in the recorded values. Wind speed is used as
an example in this section. Other variables will have different units and
different sensitivities but the principle is the same. For sub-system
checks, use the manual for specific guidance.
Analog
system
Some
systems contain "calibration" switches which are designed to test
the stability of the circuits and to provide a basis for adjustment if
changes occur. These should certainly be exercised during routine
calibrations when data loss is expected because of calibration. In the
hierarchy of calibrations, wind tunnel is first, known rate of rotation is
second, substitute frequency is third and substitute voltage is fourth. The
"calibration" switch is either third or fourth.
If
analog strip chart recorders are used, they should be treated as separate
but vital parts f the measurement system. They simply convert voltage or
current to a mark on a time scale printed on a continuous strip of paper or
composite material. The output voltage or current of the signal conditioner
must be measured with a calibrated meter during the rate of rotation
challenge. A simple transfer function, such as 10 m/s per volt, will provide
verification of the measurement circuit at the output voltage position. The
recorder can be challenged separately by inputting known voltages and
reading the mark on the scale, or by noting the mark position when the rate
of rotation and output voltage are both known. See the recorder manual
for recommendations should problems arise.
This
special concern with recorders results from the variety of problems which
analog recorders can introduce. A good measurement system can be degraded by
an inappropriate recorder selection. If resolution is inadequate to
distinguish between 1.3 m/s and 1.5 m/s, a 0.2 m/s accuracy is impossible.
If enough resolution is just barely there, changes in paper as a function of
relative humidity and changes in paper position as it passes the marking pen
and excessive pen weight on the paper can be the limit of accuracy in the
measurement. If the strip chart recorder is used only as a monitor and not
as a backup for the primary system, its accuracy is of much less importance.
The recorder from which data are recovered for archiving is the only
recorder subject to measurement accuracy specifications.
Digital
system
A
digital system may also present a variety of concerns to the calibration
method. One extreme is the digital system which counts revolutions or pulses
directly from the sensor. No signal conditioning is used. All that happens
is controlled by the software of the digital system and the capability of
its input hardware to detect sensor pulses and only sensor pulses. The same
challenge as described in 8.3.1.2 is used. The transfer function used to
change rate of rotation to m/s should be found in the digital software and
found to be the same as specified by the manufacturer or wind tunnel test.
If any difference is found between the speed calculated from the known
number of revolutions in the synchronous time period and the speed recorded
in the digital recorder, a pulse detection problem is certain. A receiving
inspection test may not uncover interference pulses which exist at the
measurement site. For solution of this type of problem, see the digital
recorder manufacturer's manual or recommendations.
A
digital data logger may present different concerns. It may be a device which
samples voltages, averages them, and transfers the average to a memory
peripheral, either at the site or at the end of a communication link.
Conversion to engineering units may occur at almost any point. The routine
calibration should look at the output voltage of a signal conditioner as a
primary point to assess accuracy of measurement. Analog to digital
conversion, averaging and transmission and storage would be expected to
degrade the measurement accuracy very little. Such functions should
contribute less than 0.05 m/s uncertainty from a voltage input to a stored
average value. If greater errors are found when comparing known rates of
rotation and known signal conditioning output voltages to stored average
wind speed values, check the data logger manual for specifications and
trouble-shooting recommendations.
8. QUALITY ASSURANCE AND QUALITY CONTROL
8.1 Instrument Procurement
8.1.1 Wind Speed
8.1.2 Wind Direction
8.1.3 Temperature and Temperature Difference
8.1.4 Dew Point Temperature
8.1.5 Precipitation
8.1.6 Pressure
8.1.7 Radiation
8.2 Installation and Acceptance Testing
8.2.1 Wind Speed
8.2.2 Wind Direction
8.2.3 Temperature and Temperature Difference
8.2.4 Dew Point Temperature
8.2.5 Precipitation
8.2.6 Pressure
8.2.7 Radiation
8.3 Routine Calibrations
8.3.1 Sensor Check
8.3.2 Signal Conditioner and Recorder Check
8.3.3 Calibration Data Logs
8.3.4 Calibration Report
8.3.5 Calibration Schedule/Frequency
8.3.6 Data Correction Based on Calibration Results
8.4 Audits
8.4.1 Audit Schedule and Frequency
8.4.2 Audit Procedure
8.4.3 Corrective Action and Reporting
8.5 Routine and Preventive Maintenance
8.5.1 Standard Operating Procedures
8.5.2 Preventive Maintenance
8.6 Data Validation and Reporting
8.6.1 Preparatory Steps
8.6.2 Levels of Validation
8.6.3 Validation Procedures
8.6.4 Schedule and Reporting
8.7 Recommendations